The CALIBRATION – SPECIALISTS for FLOWELEMENTS

External Company, but located within SEIKO Production site in Blatnize CZ.

Calibration equipment built by SEIKO.

SYSTEM PARAMETERS:

Calibration Medium: Hot water with max. 80 °C
System pressure: maximum 10,0 bar
Pressure difference: maximum 6,0 bar
Volume current: maximum 4400 m3/h
Diameter of pipe: between 80 mm and 1500 mm
Uncertainty of C: 0,10 % reachable (depends which kind of Flowelement)

Advantages:                     

  • high availability und flexibility
  • Hot water calibration (80 °C) can reach a triplicate Reynoldsnumber as by cold water (20 °C)